Skip to main content
Network Thinking Solutions Logo

CVE-2019-25719

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions.

All CVEs
CVE-2019-25719Published 6/2/2026

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions.

Vendor advisory
CVSS8.8High

Overview

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Published
6/2/2026
Last modified
7/22/2026
Assigned by
Not available
NVD analysis
Not available

Exploitation

No CISA KEV flag in the saved record

An absent KEV flag does not establish that exploitation has never occurred. Verify current intelligence before prioritizing work.

EPSS predictions and CISA SSVC assessments are not imported here. Review additional intelligence on Cloudflare

Affected products

Affected-version rules published in the advisory. Conditions and uncertain comparisons require verification.

Vendor / productAffected versionsPlatform / conditionsEvidence
Infinity Acute Care System
Dräger
≤ VG4.1.1

Verify advisory

Source
Infinity Acute Care System
Dräger
≤ VG4.0.3

Verify advisory

Source
Standalone Infinity M540 patient monitor
Dräger
≤ VG4.1.1

Verify advisory

Source

Recommended actions

Verify the product, installed version, platform, and advisory prerequisites on each system. Apply the vendor's applicable fixed release or mitigation. An affected-range boundary alone is not a universal update recommendation.

Vendor advisory links

Derived from non-aggregator references because the source record has no vendor tags. Verify the advisory before acting on it.

Severity

CVSS describes technical impact. It is separate from exploitation evidence.

Published headline score: 8.8. Score version, vector, and attribution are not available in this record. Additional score details appear when the source record is next refreshed.

Sources & references

Sources: Official CVE List, NVD and cve.now (summaries CC BY 4.0). Uses NVD data; not endorsed by NVD.