Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions.
Overview
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
- Published
- 6/2/2026
- Last modified
- 7/22/2026
- Assigned by
- Not available
- NVD analysis
- Not available
Exploitation
No CISA KEV flag in the saved record
An absent KEV flag does not establish that exploitation has never occurred. Verify current intelligence before prioritizing work.
EPSS predictions and CISA SSVC assessments are not imported here. Review additional intelligence on Cloudflare
Affected products
Affected-version rules published in the advisory. Conditions and uncertain comparisons require verification.
| Vendor / product | Affected versions | Platform / conditions | Evidence |
|---|---|---|---|
Infinity Acute Care System Dräger | < VG4.2 | Verify advisory | Source |
Infinity Acute Care System Dräger | < VG4.1.1 | Verify advisory | Source |
Infinity Acute Care System Dräger | < VG4.0.3 | Verify advisory | Source |
Standalone Infinity M540 patient monitor Dräger | < VG4.2 | Verify advisory | Source |
Standalone Infinity M540 patient monitor Dräger | < VG4.1.1 | Verify advisory | Source |
Recommended actions
Verify the product, installed version, platform, and advisory prerequisites on each system. Apply the vendor's applicable fixed release or mitigation. An affected-range boundary alone is not a universal update recommendation.
Vendor advisory links
Derived from non-aggregator references because the source record has no vendor tags. Verify the advisory before acting on it.
Severity
CVSS describes technical impact. It is separate from exploitation evidence.
Published headline score: 8.8. Score version, vector, and attribution are not available in this record. Additional score details appear when the source record is next refreshed.
Sources & references
Sources: Official CVE List, NVD and cve.now (summaries CC BY 4.0). Uses NVD data; not endorsed by NVD.