Skip to main content
Network Thinking Solutions Logo

CVE-2026-0611

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .

All CVEs
CVE-2026-0611Published 6/2/2026

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .

Vendor advisory
CVSS9.2Critical

Overview

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

Published
6/2/2026
Last modified
7/22/2026
Assigned by
Not available
NVD analysis
Not available

Exploitation

No CISA KEV flag in the saved record

An absent KEV flag does not establish that exploitation has never occurred. Verify current intelligence before prioritizing work.

EPSS predictions and CISA SSVC assessments are not imported here. Review additional intelligence on Cloudflare

Affected products

Affected-version rules published in the advisory. Conditions and uncertain comparisons require verification.

Vendor / productAffected versionsPlatform / conditionsEvidence
Sentinel
Spacelabs Healthcare
≥ 10.5.0, < 11.6.0

Verify advisory

Source

Recommended actions

Verify the product, installed version, platform, and advisory prerequisites on each system. Apply the vendor's applicable fixed release or mitigation. An affected-range boundary alone is not a universal update recommendation.

Severity

CVSS describes technical impact. It is separate from exploitation evidence.

Published headline score: 9.2. Score version, vector, and attribution are not available in this record. Additional score details appear when the source record is next refreshed.

Sources & references

Sources: Official CVE List, NVD and cve.now (summaries CC BY 4.0). Uses NVD data; not endorsed by NVD.